Masinde Muliro University of Science & Technology

Echocardiography — Quiz (20 Questions)

ECE 331: Introduction to Biomedical Engineering  |  Department of Electrical & Communication Engineering

3rd Year Undergraduate — Electrical & Communication Engineering

BLOOM'S LEVELS 1–3: REMEMBER • UNDERSTAND • APPLY

Instructions

Remember — Q1–Q7 Understand — Q8–Q14 Apply — Q15–Q20
Q1Remember

Diagnostic cardiac ultrasound typically uses which frequency range?

Q2Remember

The average speed of sound in human soft tissue used by ultrasound scanners is approximately:

Q3Remember

Echocardiography is preferred over many other cardiac imaging methods primarily because it:

Q4Remember

The piezoelectric crystal material traditionally used in ultrasound transducers is:

Q5Remember

In A-mode (Amplitude mode) display, the vertical axis represents:

Q6Remember

In colour Doppler, the BART convention means:

Q7Remember

M-mode echocardiography displays:

Q8Understand

Why do the blood-filled chambers of the heart (e.g., the LV cavity) appear black on a B-mode image?

Q9Understand

Increasing the transmit frequency improves image resolution but reduces penetration because:

Q10Understand

The factor of 2 appears in the Doppler equation fd = 2f₀v·cosθ/c because:

Q11Understand

If the ultrasound beam is perpendicular (θ = 90°) to the direction of blood flow, the measured Doppler shift is:

Q12Understand

Why does M-mode provide better temporal resolution than 2D B-mode for measuring fast-moving structures?

Q13Understand

Aliasing in pulsed-wave Doppler occurs when:

Q14Understand

The ALARA principle in diagnostic ultrasound states that exposure should be:

Q15Apply

An echo returns 200 µs after transmission. Using c = 1540 m/s, the depth of the reflector is:

Q16Apply

A scanner transmits at 3.5 MHz. The theoretical axial resolution (λ/2) is approximately:

Q17Apply

With f₀ = 2.5 MHz, blood velocity v = 1.0 m/s and θ = 0°, the Doppler shift (c = 1540 m/s) is:

Q18Apply

A PW Doppler system with f₀ = 2.5 MHz measures fd = 1.62 kHz at θ = 0°. The blood velocity is:

Q19Apply

Continuous-wave Doppler measures a peak velocity of 4 m/s across the aortic valve. Using the simplified Bernoulli equation (ΔP = 4v²), the peak pressure gradient is:

Q20Apply

A phased-array probe with element spacing d = 0.3 mm steers the beam to θ = 30°. The required progressive delay between adjacent elements (c = 1540 m/s) is approximately:

Your Score

0 / 20

Post-Test Answer Key & Explanations

Correct answers are highlighted in green. Study each explanation — the distractors are designed around common student errors.

▸ Show / Hide Full Answer Key
Q1. B  Remember

Diagnostic echocardiography uses 2–15 MHz. Below 2 MHz resolution is too poor; above ~15 MHz attenuation prevents adequate penetration of the chest wall. (Human hearing tops out at 20 kHz; 100 MHz+ would be radio frequencies.)

Q2. C  Remember

c ≈ 1540 m/s is the standard assumed speed (13 µs round-trip time per cm of depth). 340 m/s is the speed of sound in air — a common distractor.

Q3. B  Remember

Echo is non-invasive, portable, repeatable and free of ionising radiation. Resolution is good but not better than MRI, and image acquisition is highly operator-dependent.

Q4. B  Remember

PZT — lead zirconate titanate — converts electrical energy to mechanical vibration and vice versa, enabling the same crystal to transmit and receive.

Q5. B  Remember

A-mode plots echo amplitude versus depth (range) along a single beam line. It is the raw data from which all other modes are built.

Q6. A  Remember

BART: Blue = flow Away from the transducer, Red = flow Towards the transducer. Green/variance is added to encode turbulence.

Q7. C  Remember

M-mode (Motion mode) keeps one beam line fixed: depth on the vertical axis, time scrolling horizontally — hence very high temporal resolution.

Q8. C  Understand

Echo brightness depends on impedance mismatches. Blood is acoustically homogeneous, so almost no energy scatters back — the cavity is anechoic (black). Walls and valves, with large impedance steps, appear bright.

Q9. B  Understand

Attenuation in tissue is roughly proportional to frequency (≈ 0.5–1 dB/cm/MHz round trip). Higher f₀ → shorter wavelength (better resolution, λ/2) but the echo from depth is attenuated more — the classic resolution–penetration trade-off.

Q10. B  Understand

The moving red blood cell first 'receives' a Doppler-shifted wave (moving receiver), then re-radiates it as a moving source — producing a double shift, hence the factor 2.

Q11. C  Understand

fd ∝ cosθ and cos90° = 0. Flow crossing the beam at right angles is invisible to Doppler — which is why sonographers align the beam as parallel to flow as possible.

Q12. B  Understand

2D must spread its pulses over ~100 scan lines per frame (~30–60 frames/s), while M-mode fires repeatedly along ONE line (thousands of lines/s), capturing the exact extremes of motion for precise measurements.

Q13. B  Understand

PW Doppler samples at the PRF; the sampling theorem limits unambiguous shifts to ±PRF/2. Beyond that, high velocities 'wrap around' the baseline — the same aliasing you studied in signals & systems.

Q14. B  Understand

ALARA — As Low As Reasonably Achievable — directs operators to use the minimum acoustic output that gives a diagnostic image, limiting thermal and mechanical bioeffects.

Q15. B  Apply

d = c·t/2 = 1540 × 200×10⁻⁶ / 2 = 0.154 m = 15.4 cm. Remember the factor of 2 for the round trip — 30.8 cm is the classic wrong answer.

Q16. B  Apply

λ = c/f₀ = 1540 / 3.5×10⁶ ≈ 0.44 mm; axial resolution ≈ λ/2 ≈ 0.22 mm. Option 0.44 mm forgets the factor of 2; 0.11 mm is the value for 7 MHz.

Q17. C  Apply

fd = 2f₀v·cosθ/c = 2 × 2.5×10⁶ × 1.0 × 1 / 1540 ≈ 3247 Hz ≈ 3.25 kHz. (1.62 kHz would result from forgetting the factor of 2.)

Q18. B  Apply

v = fd·c/(2f₀cosθ) = 1620 × 1540 / (2 × 2.5×10⁶ × 1) = 0.50 m/s — a normal transmitral flow velocity.

Q19. D  Apply

ΔP = 4v² = 4 × 4² = 4 × 16 = 64 mmHg — a severe aortic stenosis gradient (normal aortic valve peak velocity ≈ 1 m/s → ΔP ≈ 4 mmHg).

Q20. B  Apply

Δt = d·sinθ/c = 0.3×10⁻³ × sin30° / 1540 = 0.3×10⁻³ × 0.5 / 1540 ≈ 97 ns. These nanosecond delays are generated by the system's digital beamformer.